Sony SIR-3400H Benutzerhandbuch

Seite 117

Advertising
background image

(E)

17-5

SIR-3400H / SIR-3032i / SIR-3032W

Command

Contents

IDS :

ID increment

Command format

IDS : <1>
<1> : 0=RSV, 1=ID increment
The recording unit ID is incremented each time the IDS : 1 command is issued.
Related command : IDS?

INP :

Input format setting

Command format

INP : <1><2>...<17>/[ ]
<1> : Measuring unit ID number

0=Measuring unit 1

1 to 16 CH

1=Measuring unit 1

17 to 32 CH

2=Measuring unit 2

33 to 48 CH

3=Measuring unit 2

49 to 64 CH

4=Measuring unit 3

65 to 80 CH

5=Measuring unit 3

81 to 96 CH

6=Measuring unit 4

97 to 112 CH

7=Measuring unit 4

113 to 128 CH

<2> : Input format for CH1

0=DIRECT-DC coupling
1=DIRECT-AC coupling
2=IEPE
3=TEDS

<3> : Input format for CH2

<17> : Input format for CH16
Related command : INP?

MEM :

Recording memo setting

Command format

MEM : <1><2>...<96>/[ ]
<1><2>...<96>: ASCII characters (20h to FFhex) up to maximum 96 characters
Ex.) MEM:MEASUREMENT-1 — (Annotation character string

“MEASUREMENT-1” set.)

Related command : MEM?

MOD :

Transport mode setting

Command format

MOD : <1>/[ ]
<1> : 0=STOP, 1=FWD, 2=E-E, 3=REC, 4=FWD-PAUSE, 5=REC-PAUSE,

6=FF, 7=RSV, 8=REW, 9=RSV, A=EJECT

Ex.) MOD : 1 — FWD mode
Related command : MOD?

MON :

Monitor channel selection

Command format

MON : <1><2>/[ ]
<1> : Measuring unit ID number

<2> : Channel code

0=Measuring unit 1

1 to 16 CH

0=CH 1

1=Measuring unit 1

17 to 32 CH

1=CH 2

2=Measuring unit 2

33 to 48 CH

2=CH 3

3=Measuring unit 2

49 to 64 CH

3=CH 4

4=Measuring unit 3

65 to 80 CH

4=CH 5

5=Measuring unit 3

81 to 96 CH

5=CH 6

6=Measuring unit 4

97 to 112 CH

6=CH 7

7=Measuring unit 4

113 to 128 CH

7=CH 8
8=CH 9
9=CH 10
A=CH 11
B=CH 12
C=CH 13
D=CH 14
E=CH 15
F=CH 16

Related command : MON?

.

.

.

17. Control by Telnet