Sony SIR-3400H Benutzerhandbuch

Seite 125

Advertising
background image

(E)

17-13

SIR-3400H / SIR-3032i / SIR-3032W

Command

Contents

MON?

Monitor channel setting readout

Command format

MON? /[ ]

Returned format

MON : <1><2>
<1> : Measuring unit ID number

<2> : Channel code

0=Measuring unit 1

0=CH 1

1=Measuring unit 1

1=CH 2

2=Measuring unit 2

2=CH 3

3=Measuring unit 2

3=CH 4

4=Measuring unit 3

4=CH 5

5=Measuring unit 3

5=CH 6

6=Measuring unit 4

6=CH 7

7=Measuring unit 4

7=CH 8
8=CH 9
9=CH 10
A=CH 11
B=CH 12
C=CH 13
D=CH 14
E=CH 15
F=CH 16

Related command : MON

OFS?

Input DC offset value readout

Command format

OFS? <1><2>/[ ]
<1> : 0=Recording unit setting value, 1=Recording media data
<2> : Measuring unit ID number (0=Recording unit)

Returned format

OFS : <1><2>...<32>
<1><2>

: CH 1 DC offset (2-digit hexadecimal 2’s complement display)

<3><4>

: CH 2 DC offset (2-digit hexadecimal 2’s complement display)

<31><32> : CH 16 DC offset (2-digit hexadecimal 2’s complement display)
Related command : OFS

OPL?

Output level setting readout

Command format

OPL? <1><2>/[ ]
<1> : 0=Recording unit setting value, 1=RSV
<2> : Measuring unit ID number (0=Recording unit)

Returned format

OPL : <1><2>...<16>
<1> : CH 1 output level code
<2> : CH 2 output level code

<16> : CH 16 output level code
Output level code
0=RSV, 1=RSV, 2=5 V, 3=2 V, 4=1 V, 5=RSV, 6=RSV, 7=RSV, 8=RSV,
9=RSV, A=Manual
Related command : OPL

PNL?

Panel lock status readout

Command format

PNL?/[ ]

Returned format

PNL : <1>
<1> : 0=Panel lock canceled, 1=Panel lock set
Related command : PNL

.

.

.

.

.

.