14-5. Executing Calibration (Calibration)

c
. Enable the calibration function using the DIP switches on the measuring unit.
. After the power is turned on, wait at least 15 minutes before performing calibration.
. Perform in the environment where it will actually be used.

1

Press the F1 key in the Measurement screen.
The Utility screen is displayed.

2

Use the f / F key to select “Calibration”.

3

Press the F1 key.
Calibration is performed.

<To stop calibration>

Press the F4 key. Calibration is stopped at the point where the frequency band changes.
If calibration is stopped before completion, turn the power off and then on again before using.

14. Execution of Function/Checking