Preparation, Or to correct false alarms, Notice – VACUUBRAND BVC professional /G: Instruction d´ajustement le capteur de niveau de liquide Benutzerhandbuch

Seite 12

Advertising
background image

12/20

At the factory, the level sensor is adjusted either to the 4l polypropylene bottle or to the coated 2l glass

bottle (G version) at standard conditions.

It may be necessary to adjust the sensor to compensate for changes in bottle material, bottle size or

changed ambient conditions.

Adjustment of the BVC level sensor to work properly

with a new bottle, or to correct false alarms

Pertains only to the BVC professional / BVC professional G

NOTICE

Preparation

1. Do not switch off the BVC.

2. To shut off a false alarm prior to an adjustment, simply touch the key marked “bottle change.”

The red LED will continue to flash.

3. Empty the bottle.

4. Set the correct bottle size at the touch panel.

5. Wait approx. 1 minute before adjustment.

We also recommend emptying the bottle and adjusting the BVC professional if there is a false alarm

indicating “liquid level” even though the bottle is less than 3/4 full.

If there is a false alarm in spite of an adjustment, repeat the adjustment.

If there is a false alarm in spite of

repeated adjustment, please call +49 9342 808-5500 (after sales

service), or your local VACUBRAND sales office.

Do not attach adhesive foil or anything similar to level sen-

sor or to the side of the bottle directly next to the sensor.

adhesive foil

NOTICE

Use only the original VACUUBRAND BVC bottle or the original spare bottle (see

“Accessories - Spare parts”).