The proper adjustment routine depends on two factors:

1. the type of collection bottle (2l glass bottle or 4l polypropylene bottle); and,

2. the software version of your BVC professional model.

For software versions V1.6 and higher, the version number is on the rating plate. For earlier models, the

software version is determined by the product serial number.

Note: If your BVC professional is repaired at the factory, we will update the software to the latest version,

and print the software version on the rating plate.

Selection of the adjustment routine

Signal tones

Alert tone: Repeated beep

The signal tone is a repeated beep if the bottle is more than 3/4 full, or if there is a false alarm.

Buzzer: A single buzzing sound

The signal tone is a single buzzing sound if the adjustment routine has been carried out correctly follow-

ing adjustment routines 1, 2, 3 and 4.

Bottle

Software version

Serial number

Adjustment

routine

4l PP

V1.6 and higher

not relevant

1

2l Glass

V1.6 and higher

not relevant

2

4l PP

not specified

36755501-37040401

3

2l Glass

not specified

37040501- 37040502

4

4l PP

not specified

35884001-36612112

5

2l Glass

not specified

35884201-36546103

6

Select the adjustment routine using the table below.

rating plate

VACUUBRAND GMBH + CO KG

BVC professional

2013/04

SN 37146001 / V1.6
230 V / 50/60 Hz; 0.7 A
max. 0.7/0.85 m

3

/h

150 mbar

® Alfred-Zippe-Str. 4

97877 Wertheim

Made in Germany

serial number

software version

rating plate

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: