Adjustment routine 2 – VACUUBRAND BVC professional /G: Instruction d´ajustement le capteur de niveau de liquide Benutzerhandbuch

Seite 15

Advertising
background image

15/20

Action

Reaction

Meaning

Step 1

Touch and hold “2l Glass” key.

Case 1:

The LED of the “2l Glass“ bottle

symbol flashes blue.

Adjustment is possible by

continuing to hold the “2l Glass”

key, and proceeding to step 2.

Case 2 / false alarm:

The alert tone of the false alarm

is shut off.

The LED of the “2l Glass“ bottle

symbol flashes blue.

Step 2

When the LED of the “2l Glass”

key flashes blue, touch and hold

the “bottle change” key at the

same time.

A one-time buzzer sounds.

After the buzzer, the LED of the

“bottle change“ key lights red.

Adjustment has been carried out

correctly.

Step 3

Release both “2l Glass” key and

“bottle change” key.

The LED of the “2l Glass“ key

lights blue.

The red LED of the “bottle

change” key turns off.

Adjustment finished.

Adjustment routine 2

Attention: Do not interrupt the adjustment routine once you have begun, even if

there is an alarm. Complete the entire routine, and repeat if necessary.

Bottle

Software version

Serial number

Adjustment

routine

2l Glass

V1.6 and higher

not relevant

2