Action

Reaction

Meaning

Step 1

Touch “4l PP“ key.

The LED of the „4l PP“ bottle

symbol lights blue.

Ready for adjustment.

Step 2

Release “4l PP“ key.

Step 3

Touch and hold „2l Glass“ key. The LED of the „4l PP“ bottle

symbol turns off.

The LED of the „2l Glass“ bottle

symbol lights blue.

Adjustment is possible by con-

tinuing to hold the “2l Glass” key,

and proceeding to step 4.

Step 4

When the LED of the „2l Glass“

key lights blue, touch and hold

the “bottle change” key at the

same time.

Case 1:

A one-time buzzer sounds.

After the buzzer, the LED of the

“bottle change” key lights red.

Adjustment has been carried out

correctly.

Case 2 / false alarm:

The alert tone for the false alarm

is shut off.

A one-time buzzer sounds.

After the buzzer, the LED of the

“bottle change” key lights red.

Continue with step 5 and then

repeat adjustment after 1 minute.

Step 5

Release both „2l Glass“ key and

„bottle change“ key.

The LED of the “2l Glass” key

lights blue.

The red LED of the “bottle

change” key turns off.

Adjustment finished.

Adjustment routine 4

Attention: Do not interrupt the adjustment routine once you have begun, even if

there is an alarm. Complete the entire routine, and repeat if necessary.

Bottle

Software version

Serial number

Adjustment

routine

2l Glass

not specified

37040501- 37040502

4

Advertising