Adjustment routine 6 – VACUUBRAND BVC professional /G: Instruction d´ajustement le capteur de niveau de liquide Benutzerhandbuch

Seite 19

Advertising
background image

19/20

Action

Reaction

Meaning

Step 1

Touch „4l PP“ key.

The LED of the “4l PP” bottle

symbol lights blue.

Ready for adjustment.

Step 2

Release „4l PP“ key.

Step 3

Touch and hold

“2l Glass“ key.

The LED of the “4l PP” bottle

symbol turns off.

The LED of the “2l Glass” bottle

symbol lights blue.

Adjustment is possible by

continuing to hold the “2l Glass”

key, and proceeding to step 4.

Step 4

When the LED of the “2l Glass”

key lights blue, touch and hold

the “bottle change” key at the

same time.

Case 1:

The LED of the key „bottle

change“ lights red.

Adjustment has been carried out

correctly.

Case 2 / false alarm:

The LED of the “bottle change”

key flashes red and the “false

alarm” alert is shut off. Release

the “bottle change” key and

touch a second time. The LED

of the “bottle change” key lights

red.

Continue with step 5 and then

repeat adjustment after 1 minute.

Step 5

Release both “2l Glass” key and

“bottle change” key.

The LED of the “2l Glass“ key

lights blue.

The red LED of the “bottle

change” key turns off.

Adjustment finished.

Adjustment routine 6

Attention: Do not interrupt the adjustment routine once you have begun, even if

there is an alarm. Complete the entire routine, and repeat if necessary.

Bottle

Software version

Serial number

Adjustment

routine

2l Glass

not specified

35884201-36546103

6

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: