Guidance and manufacturer’s declaration - electromagnetic immunity

The FlexScan L560T-C is intended for use in the electromagnetic environment specified below.

The customer or the user of the FlexScan L560T-C should assure that it is used in such an environment.

Immunity test

IEC60601 test level Compliance level

Electromagnetic environment -

guidance

Electrostatic

discharge (ESD)

EN61000-4-2

±6kV contact

±8kV air

±6kV contact

±8kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered

with synthetic material, the relative

humidity should be at least 30%.

Electrical fast

transient / burst

EN61000-4-4

±2kV

for power supply lines

±1kV

for input/output lines

±2kV

for power supply lines

±1kV

for input/output lines

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

EN61000-4-5

±1kV line(s) to line(s)

±2kV line(s) to earth

±1kV line(s) to line(s)

±2kV line(s) to earth

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage dips, short

interruptions and

voltage variations on

power supply input

lines

EN61000-4-11

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycles

70% U

T

(30% dip in

U

T

) for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5sec

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycles

70% U

T

(30% dip in

U

T

) for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5sec

Mains power quality should be

that of a typical commercial or

hospital environment. If the user

of the FlexScan L560T-C requires

continued operation during

power mains interruptions, it is

recommended that the FlexScan

L560T-C be powered from an

uninterruptible power supply or a

battery.

Power frequency

(50/60Hz)

magnetic field

EN61000-4-8

3A/m

3A/m

Power frequency magnetic fields

should be at levels characteristic

of a typical location in a

typical commercial or hospital

environment.

NOTE

U

T

is the a.c. mains voltage prior to application of the test level.