Set receive to transmit delay, Report number of floats, Report receive to transmit delay – VEGA VEGABAR 81 Modbus and Levelmaster protocol - Operating Instructions Benutzerhandbuch

Seite 74

Advertising
background image

74

10 Anhang

VEGABAR 81 • Modbus- und Levelmaster-Protokoll

46293-DE-131030

Response:

Parameter

Length

Code/Data

Set Baud Rate

11 characters ASCII

Beispiel: U01B9600E71
Gerät an Adresse 1 ändern zu Baudrate 9600, Parität even, 7 Datenbits, 1 Stoppbit

Set Receive to Transmit Delay

Request:

Parameter

Length

Code/Data

Set Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 up to 250),

default = 127 ms

Response:

Parameter

Length

Code/Data

Set Receive to Transmit Delay

6 characters ASCII

UuuROK

Report Number of Floats

Request:

Parameter

Length

Code/Data

Set Receive to Transmit Delay

4 characters ASCII

UuuF

Response:

Parameter

Length

Code/Data

Set Receive to Transmit Delay

5 characters ASCII

UuuFn
n = number of measurement values

(0, 1 or 2)

Report Receive to Transmit Delay

Request:

Parameter

Length

Code/Data

Report Receive to Transmit Delay

4 characters ASCII

UuuR

Response:

Parameter

Length

Code/Data

Report Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 up to 250),

default = 127 ms

Advertising