VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Approval according to LPR radio standard Benutzerhandbuch

Seite 82

Advertising
background image

P

arameter

L

ength

C

ode/Data

Assign Unit Number

6

characters ASCII

UuuNOK
uu = new Address

S

et number of Floats

R

equest:

P

arameter

L

ength

C

ode/Data

Set number of Floats

5

characters ASCII

UuuFn

R

esponse:

P

arameter

L

ength

C

ode/Data

Set number of Floats

6

characters ASCII

UuuFOK

Wird die Anzahl auf 0 gesetzt, wird kein Füllstand mehr zurückgemeldet

S

et Baud Rate

R

equest:

P

arameter

L

ength

C

ode/Data

Set Baud Rate

8 (12)

characters ASCII

UuuBbbbb[b][pds]
Bbbbb[b] = 1200, 9600 (default)
pds = parity, data length, stop bit
(

optional)

parity: none = 81, even = 71 (de-
fault), odd = 71

R

esponse:

P

arameter

L

ength

C

ode/Data

Set Baud Rate

11

characters ASCII

Beispiel: U01B9600E71

Gerät an Adresse 1 ändern zu Baudrate 9600, Parität even, 7 Datenbits, 1 Stoppbit

S

et Receive to Transmit Delay

R

equest:

82

VEGAPULS 62 • Zulassung nach LPR-Funkrichtlinie

10 A

nhang

41722

-DE