3 what to do if, What to do if, 2cleaning – Xylem LF313T_LF313T-ID Benutzerhandbuch

Seite 13: 3what to do if

Advertising
background image

LF313T

Cleaning

13

ba75522e03

10/2009

2

Cleaning

Caution

Before cleaning, disconnect the sensor from the instrument.

Cleaning

A thorough cleaning is particularly recommended for measurements of low
conductivities.

After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.

Aging of the

conductivity

measuring cell

Normally, the conductivity measuring cell does not age. Special measuring
media (e.g. strong acids and lyes, organic solvents) or too high temperatures
shorten the operational lifetime considerably or damage the measuring cell.
The warranty does not cover cases where such conditions cause failure or
mechanical damage.

Disposal

We recommend to dispose of the conductivity cell as electronic waste.

3

What to do if...

Contamination

Cleaning procedure

Lime sediments

Immerse in acetic acid for 5 minutes
(volume share = 10 %)

Fat/oil

Clean with warm water that contains
washing-up liquid

Error symptom

Cause

Remedy

No temperature or
conductivity display

– No connection between

measuring instrument and sensor

– Cable defective

– Check connection between

measuring instrument and sensor

Measurement delivers
implausible conductivity
values

– Incorrect cell constant adjusted at

the measuring instrument

– Measuring range exceeded
– Contamination in the area of the

electrodes

– Electrodes damaged

– Check / correct the cell constant

– Observe the application range
– Clean the sensor (see section 2).

– Return the sensor

Incorrect temperature
display

– The temperature sensor was not

immersed deep enough in the
measuring solution

– Temperature sensor defective

– Observe the minimum immersion

depth

– Return the sensor