Aging, Maintenance – Xylem ISE Electrodes Cu 60 Benutzerhandbuch

Seite 19

Advertising
background image

19

Aging

Please note that every electrode undergoes a natural
aging process. The response time increases and the
slope decreases with the age of the electrode. The
following factors shorten the lifetime considerably:

Incorrect storing

Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of
interfering ions)

High temperatures

High changes in temperature

The warranty does not cover damage caused by
measuring conditions and mechanical damage.

Maintenance

For combination electrodes: replenish used reference
electrolyte solution.

In case of increased response times of the electrode
perform the following activities:

- Polish the membrane surface using an SF/K
polishing strip. To do so, moisten a polishing trip
with deionized water, and with circular movements
polish the membrane surface for approx. 30
seconds.

- Subsequently place the electrode into diluted
standard solution for approx. 5 minutes.




Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: