Storage

Combination

electrode CA 60

Between two measurements, place the
combination

electrode into diluted standard

solution.

Overnight up to one week: Push the closing ring
upwards to close the refilling opening. Then rinse
the combination electrode with deionized water,
blot it dry using a clean paper towel and store it
in a dry place.

For more than a week: Remove the electrolyte
solution from the reference electrode and rinse
the combination electrode with deionized water,
blot it dry using a clean paper towel and put on
the protection cap. Store the combination
electrode in a dry place.

Halfcell Ca 1100 A

• Between two measurements, place the electrode

into diluted standard solution.

• For storing overnight or longer, rinse the

electrode with deionized water, blot it dry using a
clean paper towel and put on the protection cap
(closing cap of the glass tube). Store the
electrode in a dry place.