Aging

Please note that every electrode undergoes a natural

aging process. The response time increases and the
slope decreases with the age of the electrode. The
following factors shorten the lifetime considerably:

• Incorrect storing
• Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of

interfering ions)

• High temperatures
• High changes in temperature

The warranty does not cover damage caused by
measuring conditions and mechanical damage.

Maintenance

• For combination electrodes: replenish used reference

electrolyte solution.

In case of increased response times of the electrode

perform the following activities:

- Polish the membrane surface using an SF/K
polishing strip. To do so, moisten a polishing trip
with deionized water, and with circular movements
polish the membrane surface for approx. 30
seconds.

- Subsequently place the electrode into diluted
standard solution for approx. 5 minutes.