The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

3. Das Symbol für die Blindprobenmessung wird

angezeigt. Öffnen Sie den Gerätedeckel und
pipettieren Sie ein Aliquot von 2,0 µl der
Matrix-Blindprobe auf den unteren Lesekopf.

Schließen Sie den Gerätedeckel und drücken
Sie auf die Taste neben dem Häkchen-Symbol,
um die Leerwertmessung durchzuführen.

4. Das Gerät nimmt fünf Messungen bei 260 und

330 nm und den Schichtdicken 0,5 und 0,2 mm vor.

6. Das Symbol für die Kalibrierlösungsmessung

wird ange zeigt. Öffnen Sie den Gerätedeckel
und verwenden Sie ein fusselfreies Tuch, um
die Leseköpfe zu reinigen.

7. Pipettieren Sie ein Aliquot von 2,0 µl der

Kalibrierlösung auf den unteren Lesekopf.
Schließen Sie den Gerätedeckel und drücken Sie
auf die Taste neben dem Häkchen-Symbol, um
die Messung der Kalibrierlösung durchzuführen.

8. Das Gerät nimmt fünf Messungen bei 260 und

330 nm und den Schichtdicken 0,5 und 0,2 mm vor.

10. Das Symbol für die Kalibrierlösungsmessung

wird ein zweites Mal angezeigt. Öffnen Sie den
Gerätedeckel und verwenden Sie ein fusselfreies
Tuch, um die Leseköpfe zu reinigen.

11. Pipettieren Sie ein Aliquot von 2,0 µl der

Kalibrierlösung auf den unteren Lesekopf.
Schließen Sie den Gerätedeckel und drücken Sie
auf die Taste neben dem Häkchen-Symbol, um
die Messung der Kalibrierlösung durchzuführen.

5. Wenn die gemessenen Werte innerhalb der erforderlichen Toleranzen liegen, wird das Symbol für

„Test bestanden“ angezeigt. Drücken Sie auf die Taste neben dem Häkchen-Symbol, um fortzufahren,
oder auf die Taste neben dem Kreuz-Symbol, um den Kalibriervorgang abzubrechen.

9. Wenn die gemessenen Werte innerhalb der erforderlichen Toleranzen liegen, wird das Symbol für

„Test bestanden“ angezeigt. Drücken Sie auf die Taste neben dem Häkchen-Symbol, um fortzufahren,
oder auf die Taste neben dem Kreuz-Symbol, um den Kalibriervorgang abzubrechen.

12. Das Gerät nimmt fünf Messungen bei 260 und 330 nm und den Schichtdicken 0,5 und 0,2 mm vor.

13. Wenn die gemessenen Werte innerhalb der erforderlichen Toleranzen liegen, wird das Symbol

für „Test bestanden“ angezeigt und das Gerät berechnet die neuen Kalibrierfaktoren für die
Schichtdicke. Drücken Sie auf die Taste neben dem Häkchen-Symbol, um fortzufahren, oder auf
die Taste neben dem Kreuz-Symbol, um den Kalibriervorgang abzubrechen.

14. Die neuen Werte für den Kalibrierfaktor müssen nun durch erneute Messung der Luft-, Blind- und

Standardwerte überprüft werden.