Electrical connections – Vaisala GMW20 Benutzerhandbuch

Seite 6

Advertising
background image

User’s Guide_____________________________________________________ GMW20

4

Electrical Connections

See the requirements for the power supply in the Technical

Data section.

1.

Connect the nominal 24 V supply on the PCB between the

terminals + and -. Connect the common wire to terminal 0

and the other wire either to terminal V (voltage output) or

to the terminal mA (current output).

2.

Choose the current output with the jumper 0/4 mA

– 4 ... 20mA: jumper shorts the pins (default)

– 0 ... 20 mA: disconnect (do not discard) the jumper.

3.

If the unit has an optional accessory (relay, display and

relay, LonWorks interface, or temperature module), follow

the procedure described in the applicable manuals before

repositioning the cover.

4.

Reposition the cover.

Current

output

jumper

Power supply
terminals

Output signal
terminals

Figure 2

Jumper and Terminals

Connecting power wires to the output terminal can
seriously damage the product.

Advertising