Transporting, Installing, Transferring and Disposing of the Centrifuge 5418 R

1

Entwurf

Transporting, Installing, Transferring and Disposing of the Centrifuge 5418 R

Instructions for use

These instructions for use complement the operating manual for the Centrifuge 5418 R and do

not replace it. You should therefore also read the operating manual before using the centrifuge

for the first time. The current version of the operating manual can be found in the Internet at

www.eppendorf.com.

1 Transport

Only transport the device in the original packaging.

The centrifuge must always be transported by two people.

Use a transport aid for transporting over longer distances.

2 Installation

The weight of the centrifuge is 22 kg. The centrifuge must always be transported by two people.

2.1 Open the carton and lift the device

2.2 Place device on a stable, horizontal and non-resonant lab bench.

1. Cut the adhesive tape.
2. Open all 4 flaps of the carton.
3. Remove accessories.
4. Reach your hands under the device and lift

the centrifuge from the carton with another

person.

Remove front and back transport protection

pads.

Remove the plastic sleeve

Observe a minimum distance of 30 cm to

adjoining devices at the sides and from the

rear side to the wall.

Install the device in a well-ventilated location

which is protected from direct sunlight to

prevent it from overheating.