2 installation, Connections and dimensions, type 707011, Installation – JUMO 707010 dTRANS T01 (T) HART - 2-Wire Transmitter Operating Manual Benutzerhandbuch
Seite 23
Advertising
5
2 Installation
Connections and dimensions, type 707011/...
Connection for
Terminals
Supply
10 — 35V DC
or
current output
4 — 20mA
+1
-2
R
B
= burden resistance
U
b
= supply voltage
Analog inputs
Thermocouple
+4
-6
Resistance
thermometer
in 2-wire
circuit
3
6
R
L
≤
11
Ω
R
L
= lead resistance
per conductor
Resistance
thermometer
in 3-wire
circuit
3
5
6
R
L
≤
11
Ω
R
L
= lead resistance
per conductor
Resistance
thermometer
in 4-wire
circuit
3
4
5
6
R
L
≤
11
Ω
R
L
= lead resistance
per conductor
R
B
U
b
10V
–
22mA
--------------------
=