8 Setup program

The setup program is available for configuring the transmitter from a

PC. Connection is via the HART

®

modem outside the hazardous

(Ex) area.

Configurable parameters:

- TAG number (8 digits),

16 characters of description and
32 characters of message

- sensor type, linearization

- connection circuit (2-/3-/4-wire circuit)

- lead resistance with 2-wire circuit

- unit

- internal cold junction or

external cold junction (fixed value)

- custom linearization

- range limits

- output signal rising/falling (reversed)

- digital filter

- response to probe break/short-circuit

- measurement offset

- min. / max. value

- current simulation