Setup interface, Fine calibration – JUMO 707010 dTRANS T01 Junior - 2-Wire Transmitter Operating Manual Benutzerhandbuch

Seite 17

Advertising
background image

7

Setup interface

The setup interface is available for configuring the transmitter from
a PC. Connection is via the PC interface with USB/TTL converter
and adapter.

Configurable parameters:

• TAG number (10 characters)

• sensor type

• connection circuit (2-/3-/4-wire circuit)

• external or internal cold junction

• range limits

• output signal rising/falling (reversed)

• digital filter

• response on probe break/short-circuit

• recalibration (fine calibration)

• lead resistance for 2-wire circuit

If no supply unit (supply isolator) is available, the 2-wire transmitter
can also be configured using a 9V block battery as a power source.

Please notice that the dTRANS T01 Junior doesn‘t support
custom linearization!

Fine calibration

Fine calibration means the adjustment of the output signal. The sig-
nal can be adjusted within ± 5% of 20mA full scale. Fine calibration
can be performed through the setup program. Values for 4 mA
(zero) and 20 mA (full scale) and offset can be calibrated separately
using the setup program.

Advertising