Gossen MAVOLUX 5032 C BASE Benutzerhandbuch

Seite 45

Advertising
background image

GOSSEN Foto- und Lichtmesstechnik GmbH

21

Characteristics MAVOLUX 5032C BASE

Measuring Quantity

Measuring Range

Measuring Range

Resolution

Resolution

in Lux (lx)

in footcandle (fc)

in lx

in fc

Illuminance

I

0.1... 199.9

0.01... 19.99

0.1

0.01

II

1... 1 999

0.1... 199.9

1

0.1

III

10... 19 900

1... 1 999

10

1

IV

100... 199 000

10... 19 990

100

10


Most Important Error Limits MAVOLUX 5032 C BASE

Characteristics

Admissible Error according to

DIN 5032 Klasse C

Typical Error

MAVOLUX 5032C BASE

V(

)-Adaptation (f

1

)

9%

7,5%

True Cosine Evaluation (f

2

)

6%

2,0%

Linearity (f

3

)

5%

1,5%

Adjustment Error (f

11

)

2%

1,0%

Overall Error (f

ges

)

20%

15,0%