DABx_MP4M_BA_V2_3m.doc

25/28

Baumer Electric AG

19.12.07/rut

Frauenfeld, Switzerland

4.10 Error case identification


If the measuring value is out of the
measuring range, it is indicated as described:
The error case identification allows the
control of the internal function of the amplifier
and the function of the sensor. Following
Failure can be detected:

Interrupted connection between sensor
and amplifier

Measuring value out of measurable
range

Internal functional error of amplifier

No supply voltage on the amplifier

Bridge to completion resistor not existing

1. The Error-LED changes from ’green’ to

‘red’

2. The switching output ‘Error’ has been

activated (indicates ‘Low’)

3. Representation on the LCD

SIG OFVL

PEAK 8.04V

SIG UFVL

PEAK 8.04V

If the measuring value is in the measurable
range again, the LED lights green, the
switching output is deactivated and the LCD
shows the correct value.


Note
The error output switch is default set on
‘High’. In error case the switch switches to
‘Low’!

Comment
Should the amplifier remain in error status,
follow following procedure:
Check if the sensor is fully functional
Check if the sensor is connected correctly
Tare the amplifier by activating the Reset
function.

If the amplifier still remains in error status ,
please get in contact with our representative
in your area.

Input signal

Error switching output

Upper range
limit

lower range
limit

High

Low

E

rr

o

r

c

a

s

e

E

rr

o

r

c

a

s

e

P

o

w

e

r

u

p

4.11 Start procedure

After starting the following indications appear:

1. All segments
2. ‘Baumer’
3. All segments
4. ‘DABx’
5. All segments
6. ‘Program Version’

DABx_MP4M_BA_V2_3m.doc

26/28

Baumer Electric AG

19.12.07/rut

Frauenfeld, Switzerland

5 Technical data

5.1 Electrical data

DABU

DABI

General

Voltage supply range

15 – 33 VDC

15 – 33 VDC

Supply current

< 120mA without bridge excitation

< 120mA without bridge excitation

Reset- and mode – Input

(galvanic disconnected)

5 – 33 VDC

<1 VDC

active

inactive

5 – 33 VDC

<1 VDC

active

inactive

Warm-up time

min. 10 min

min. 10 min

Input signal

Bridge excitation

5.1V

±

2%, < 15 mA

5.1V

±

2%, < 15 mA

S/G bridge resistance

350

350

Bridge completion resistors

350 Ohm

350 Ohm

Signal input range

According part number key

According part number key

Sample rate

< 1 ms

< 1 ms

Analog output

Output signal

+/- 10 VDC

Load resistance > 10kOhm

4...20mA

apparent ohmic resistance <500Ohm

Characteristic curve deviation

<

±

0.08 % FSR ( <

±

16mV )

<

±

0.15 % FSR ( <

±

24

µ

A )

Resolution Vout / Iout ( 12bit )

< 0.035 % FSR ( 7 mV )

< 0.07 % FSR ( 11

µ

A )

Resolution Vout2 / Iout2 ( 12bit )

< 0.035 % FSR ( 7 mV )

< 0.07 % FSR ( 11

µ

A )

Reset

Zero point reset activ

< ± 10 mV

< ± 10

µ

A

Tare range (reset)

±

6 mV / V

±

6 mV / V

Reset-pulse

> 1 ms

> 1 ms

Reset- settle time

> 5 ms after reset-pulse

> 5 ms after reset-pulse

Reset / operate offset

< ± 10 mV

< ± 15

µ

A

Software

Software version

see chapter 4.11

Display

Display updates

2 per second

2 per second

Resolution

10 mV

10

µ

A

Switching output

switching hysteresis threshold

< 0.25 % FSR ( 50 mV )

< 0.5 % FSR ( 80

µ

A )

Max. load limit switches

50 mA

50 mA

5.2 Mechanical data

Control and sensor connection

13 pin spring block- 0,5 ... 1,5 mm

2

( 20AWG to 16 AWG )

Shield connection

2 pin terminal block

Enclosure

Aluminium / Plastic