2

English

Welch Allyn Spot Vital Signs

5.

Press the Mode (

) button once to save the change. Repeat steps 4 and 5 to set

the day, month, year, hour, and minute values.

6.

Press the Power (

) button to exit the Internal Configuration Mode.

Configuration Menu Options

Setting

Description

Blood Pressure Calibration
Displays “Cal”

Prepares the Spot Vital Signs for calibration. Only qualified personnel should verify the
Spot Vital Signs blood pressure calibration.

Inflation Pressure Preset Level
Displays “PrP”

120, 140, 160, 180, 200, 240, 280 mmHg. Factory default is 160 mmHg.

Pressure Preset Level
Displays “PrP“

On or off. Disables or enables the front panel Pressure Preset button.

Backlight
Displays “BLT”

On or off.

Mean Arterial Pressure
Displays “MAP”

On or off.

Date/Time

Changes or updates the actual date and time.

Temperature Scale
Displays “TMP MOD”

Fahrenheit (°F) or Celsius (°C) Normal Mode / Fahrenheit (°F) or Celsius (°C) Monitor
Mode

Blood Pressure Units
Displays “BP”

mmHg or kPa.

Battery Readings
Displays “BAT”

Displays the total battery voltage.

Battery Life
Displays “LFE”

Total number unit measurements. Displayed information only; operator cannot change.