Set up the device
After recharging a dead battery or after disconnecting the battery for a few minutes, you
must program the date and time screen.
To Enter the Internal Configuration Mode:
1.
Turn the Spot LXi off (
).
2.
Press and hold the Select (
) and Power (
) buttons for 5 seconds. The device
enters the Internal Configuration Mode and the Configuration Menu screen appears
on the display.
3.
Use the Navigation (
) buttons to move through the menu options and press the
Select (
) button to access the options or accept a change.
4.
Press the Power (
) button to exit the Internal Configuration Mode.
Select
Configuration Menu
Version
Numbers
Manual
Parameters
External
Devices
Save Readings
Buttons
Lockout
Defaults
Event Log
Battery
Date/Time
Blood Pressure
Temperature
Location
Identifier
Contrast