haze-gard plus




Calibration Procedure with Taber Abrasion Holder



• Calibrate the instrument without the Taber Abrasion Holder.

• Install the Taber Abrasion Holder in front of the sphere entrance.

• Calibrate the instrument once more, but only dark- and bright calibration.

When you are asked to calibrate Clarity, please press the ”set-up“ button.

• Perform measurements without a sample in Haze, Transmission and Clarity.

The result should be: H = 0.00, T = 100 and C = 100.


If the Taber Abrasion Holder is removed, calibrate the instrument again.



Note:

When working with the Taber Abrasion Holder, only measure Haze or Transmission.
Clarity measurements are not possible with the holder due to changed
optical conditons.