ALTANA Esp-10 Benutzerhandbuch

Seite 24

Advertising
background image

• Disconnect the instrument from

the compressed-air supply.

• Remove the 4 cross-head rear

panel screws.

• Slightly tilt the raer panel with

holder backwards and remove it
by carefully pulling it up and
guiding it beyound the
counterweight.

• Remove the counterweight from

the panel support.

• Loosen the 2 screws below the

panel support using a metric 2.5
mm hexagon socket screw key.

• Carefully raise and remove the

panel support so that the spring
and the impact tool lie bare on the
guide tube.

• Carefully remove the impact tool

and the spring.

• Hold the instrument upside-down

in order to remove the ball from
the guide tube.

7. Replacing the

Impact Tool

• Check the ball for any defects.

Then put the ball back in the
guide tube or replace it
respectively.

• Put the spring in place.

• Check the impact tool for any

defects and replace it if
necessary.

• Put the impact tool in place.

• Re-assemble all other parts in

reverse order.