3 cocl, And so, Generator (see picture 3) – Compur Gas Generator Benutzerhandbuch

Seite 13

Advertising
background image

4.2.3
COCl

2

and SO

2

Generator (see Picture 3)


1 Unplug the external battery (COCl

2

only).

2 Loosen the 4 screws on the backside of the instrument and separate the two halves. (Use

caution as the LED’s are attached to the top case)

3 Remove the battery.
4 Push back the 2 isolation tubes and unsolder the generator cell from the wires.
5 Unscrew the clamp (KS) and push the generator cell backwards. Remove old generator

cell.

6 Insert the new generator cell and fasten the clamp.
7 Solder the wires to the new generator cell (polarity of no importance) and push the

isolation tubes back over the connections.

8 Insert the battery.
9 Close the housing and tighten the screws in the back. Reconnect the external battery pack.


Note:

) Whenever you have done maintenance work on a gas generator, test the function with a
working gas detector. The detector should give an alarm within 10 seconds.





























13

Advertising