Function of the analog i/o interface – VACUUBRAND Digital-I/O-Modul Benutzerhandbuch

Seite 15

Advertising
background image

Seite/page 15 / 19

Das Dokument darf nur vollständig und unverändert verwendet und weitergegeben werden. Es liegt in der Verantwortung des

Anwenders, die Gültigkeit dieses Dokumentes bezüglich seines Produktes sicher zu stellen. BA-Nr.: 999221 / 27/05/2009

Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999221 / 27/05/2009

Function of the analog I/O interface

Factory-set the analog I/O interface is configured for vacuum output as analog voltage 0-10 V, e. g. to be

conneted to a pen recorder. The input of the analog I/O interface is not used in this configuration.

If the interface is active (controller / measurement device is switched on) the LED lights green. At a pres-

sure > 1000 mbar the LED flashes green.

If the interface is inactive (controller / measurement device shut off) a voltage of -0.3 V can be measured

at the output.

Vacuum output as analog voltage 0-10 V (in combination with CVC 3000 / VNC 2 / DCP 3000)

Controller / measuring gauge with internal pressure transducer or external VACUUBRAND vacuum

sensor VSK 3000: At the output of the analog I/O interface a DC voltage of 0 - 10 V is sent. The output

voltage is linear to the vacuum. This corresponds to a pressure of 0 - 1000 mbar or 0 - 750 Torr ( 0 VB

corresponds to 0 mbar / 0 Torr, 10 V corresponds to 1000 mbar / 750 Torr).

Controller / measuring gauge with external VACUUBRAND vacuum sensor VSP 3000: At the output of

the module a DC voltage of 0.5 - 9.5 V is sent. The output voltage is logarithmic to the vacuum from

0.5 - 9.5 V with a scaling of 1.5 V per decade (totally 6 decades from 1 x 10

-3

to 1 x 10

3

mbar or 7.5 x

10

-4

to 7.5 x 10

2

Torr).

Set vacuum input and and actual vacuum output as analog voltage 0-10 V (in combination with

CVC 3000 / VNC 2)

If using a controller in operation mode „Vac control“ the analog input can be used for vacuum setting.

0 - 10 V correspond to 0 - 1000 mbar. The analog output of the analog I/O interface sends the actual

(measured) pressure value.

Motor speed output as analog voltage 0-10 V (in combination with CVC 3000)

The output of the analog I/O interface sends the actual motor speed as signal 0 -10 V (corresponds to

0 - 100 % of the motor speed).

Motor speed input and output as analog voltage 0-10 V (in combination with CVC 3000)

If using the controller in operation mode “Pump down” the analog input can be used for motor speed

setting. 0 - 10 V correspond to 0 - 100% of the motor speed.The analog output sends the actual (meas-

ured) motor speed.

If setting the vacuum or the motor speed via interface settings at the controller or via controller interface

are without function.

Note:

Modules with current input 0 - 5 V and / or output or current input 4 - 20 mA are available on request,

current output is not available.

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: