Secondary_value_1

Calibration

Lowest Point

Calibration

Highest Point

Sensor_value

cal_level_lo

cal_level_hi

Calibration
Lowest Point

Calibration
Highest Point

Sensor reference plane

cal_level_lo [%]

cal_level_hi [%]

Abb. 27: Abgleich VEGASON 66

Parameterliste für Device revision 3.0

Die folgende Liste enthält die wichtigsten Parameter und ihre Bedeutung:

l

primary_value

-

This is the process value after adjustment and Linearization with the status of the

transducer block

l

primary_value_unit

-

Unit code of 'Primary_value'

l

secondary_value_1

-

Value after min/max-adjustment (level + level offset). Selected as input to AIFB by

setting 'Channel' = 2. Unit derives from 'Secondary_value_1_unit'

l

secondary_value_1_unit

-

Unit code of 'Secondary_value_1'

l

secondary_value_2

-

Sensor value + sensor offset. Selected as input to AIFB by setting 'Channel' = 3. Unit

derives from 'Secondary_value_2_unit'

l

secondary_value_2_unit

Anhang

VEGASON 66 - Foundation Fieldbus

55