Response times

The response time depends on the concentration range.

It is:

several seconds with high concentrations,

several minutes near the detection limit.

The measured value is stable if the value does not change
by more than 0.1 mV within 30 seconds.

Interferences

Cupric, mercury and silver ions poison the membrane.

Iron and cadmium ions poison the membrane if their
concentrations are higher than that of the lead
concentration.

Aging

Please note that every electrode undergoes a natural aging
process. The response time increases and the slope
decreases with the age of the electrode. The following
factors shorten the lifetime considerably:

• Incorrect storing
• Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of
interfering ions)

• High temperatures
• High changes in temperature

The warranty does not cover damage caused by measuring
conditions and mechanical damage.