Response times

The response time depends on the concentration range.

It is:

•

several seconds with high concentrations,

•

several minutes near the detection limit.

The measured value is stable if the value does not change
by more than 0.1 mV within 30 seconds.

Interferences

•

Cupric, mercury and silver ions poison the membrane.

•

Iron and cadmium ions poison the membrane if their
concentrations are higher than that of the lead
concentration.

Aging

Please note that every electrode undergoes a natural aging
process. The response time increases and the slope
decreases with the age of the electrode. The following
factors shorten the lifetime considerably:

• Incorrect storing
• Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of
interfering ions)

• High temperatures
• High changes in temperature

The warranty does not cover damage caused by measuring
conditions and mechanical damage.