concerning sample preparation and measuring
procedures, SI Analytics provides a large number of
application reports for various applications.

Response times

The response time depends on the concentration range.
It is:

several seconds with high concentrations,

several minutes near the detection limit.

The measured value is stable if the value does not
change by more than 0.1 mV within 30 seconds.

Interferences

• Complexes with acetate, citrate, ammonia, amino

acids, EDTA etc.

• Mercury and silver ions poison the membrane.

• The membrane is photosensitive concerning the

setting of the potential.

• Interfering ions: no disturbances with the following

conditions: