Error diagnostics
Cause
Remedy
Measured value
unstable
Inner junctions not sufficiently
wetted with bridge electrolyte
Fill up bridge electrolyte until the
inner junctions are covered with
bridge electrolyte
Inner junctions encrusted
Leave the bridge electrolyte
to react on the inner junctions
for some hours until the
crusts have dissolved
Cable broken
Exchange cable
Slope too low
Membrane surface contaminated
Polish the membrane surface
with polishing strip S/FK (see
MAINTENANCE)
Conditioning time too short
Extend conditioning time
Standard solutions too old
Use new standard solutions
Inner junctions
encrusted
Leave the bridge electrolyte
to react on the inner junctions
for some hours until the
crusts have dissolved.
Electrode defective
Exchange electrode