Response times

The response time depends on the concentration range.

It is:

•

several seconds with high concentrations,

•

several minutes near the detection limit.

The measured value is stable if the value does not

change by more than 0.1 mV within 30 seconds.

Interferences

•

pH values < 5 or > 7

Aging

Please note that every electrode undergoes a natural
aging process. The response time increases and the
slope decreases with the age of the electrode. The
following factors shorten the lifetime considerably:

• Incorrect storing
• Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of
interfering ions)

• High temperatures
• High changes in temperature

The warranty does not cover damage caused by
measuring conditions and mechanical damage.