VACUUBRAND Kit EK 500 Benutzerhandbuch
Seite 13

Seite 13 von 16 // page 13 of 16
Das Dokument darf nur vollständig und unverändert verwendet und weitergegeben werden. Es liegt in der Verantwortung des
Anwenders, die Gültigkeit dieses Dokumentes bezüglich seines Produktes sicher zu stellen. 999194 // 05/02/2008
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 999194 // 05/02/2008
Assemble handle.
Use open ended wrench to reconnect hose to elbow fitting.
Tighten union nuts first by hand and then tighten one full turn
using open ended wrench.
Position housing cover.
☞ Move housing cover slightly to make sure that the head
covers are correctly positioned.
Screw in six socket head screws fixing housing cover cross-
wise first slightly, then tighten.
☞ Do not tighten until head cover is in contact with housing,
max. torque 6 Nm.
Position hose clip at the inlet hose of the exhaust waste va-
pour condenser.
Connect the hose to the outlet of the pump.
Close hose clip using a flat pliers.