The vacuum controller CVC 3000 detects the connected VMS-B-Module automatically and is configu-

rated accordingly. If no vacuum is required (or after a preset delay) the controller sends a signal to the

VMS-B-Module and the module switches off a connected vacuum pump. If two controllers are con-

nected the pump is switched off if both controllers have sent the signal that no vacuum is still required.

Switching on the pump takes place analogously and fully automatic.

Attention: VARIO pumps and pumping units are not switched by a VMS-B-Module.

In combination with pumping units PC 5xx NT or PC 6xx NT a 230V solenoid operated valve can also

be switched, however the pump is then no longer switched by the VMS-B-Module.

To switch a 230V solenoid operated valve the VMS-B-Module has to be configured approriately, see

section “Reconfiguring the VMS-B-Module”.

Reconfiguring the VMS-B-Module

(e. g. to switch a 230V / 50 Hz coolant, venting or in-line valve)

Configuration is only possible with CVC 3000 from software version V 1.30.

Reconfiguring the VACUU•BUS address of the VMS-B-Module

1. Switch the vacuum controller CVC 3000 off.

2. Preferably connect only the VMS-B-Module to be reconfigured.

3. Keep key ”VENT” pressed and switch the CVC 3000 on (key ”ON/OFF”).

4. Select the menu item ”Vacuubus” by turning and pressing the selection knob.

5. The first component detected by the VACUU•BUS is displayed. If necessary, select the component

to be reconfigured by turning and pressing the selection knob.

6. Keep key ”MODE” pressed and press selection knob additionally. The inverse display of the compo-

nent is replaced by a frame.

7. Turn the selection knob until the address, to which you want to reconfigure the VMS-B-Module (e.g.,

Isol.v. 1, Waterv. 2, ...), is displayed.

8. Press the selection knob. If the configuration was successful, the menu is quit and the CVC 3000

restarts.

A configuration of more than one components to the same VACUU•BUS address will result in error con-

ditions at the CVC 3000 and must therefore be avoided!

Further information concerning the configuration on demand.