Technical Data
We reserve the right for technical modification without prior notice!
Replacing the device fuse
Switch off the device and disconnect it from mains before replacing the fuse. After
disconnecting from mains wait two minutes to allow the capacitors to discharge.
Identify and eliminate the cause of failure before switching on the device again.
➨
Keep the snap-fit squeezed and pull the fuse holder out.
➨
The fuse holder contains two fuses of the same type. Replace the defective fuse
by a fuse of the same type (see ”Technical data”).
➨
Insert the fuse holder into the device until it snaps.
fuse holder
Type
VMS-B-Module
Mains connection
100-230 V, 50-60 Hz
Max. power / current: 24V /VACUU•BUS
TM
1.1A
Nominal current
0.8 A
Degree of protection IP
20
Max. permitted ambient temperature during operation
+10°C to +40°C
Max. permitted ambient temperature during storage
-10°C to +60°C
Permitted relative atmospheric moisture during operation
(no condensation)
30% to 85%
Max. permitted power draw at connection vacuum pump for
ohmic (inductive) load
8(4) A bei 250 V~ / 8(6) A bei 125 V~
Fuse
2 x 250 V
Dimensions
179 mm x 108 mm x 61 mm
Mass
0.7 kg