Specification

General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications
apply for the temperature range 18°C to 28°C after 1 hour warm-up with no load and calibration at
23°C. Typical specifications are determined by design and are not guaranteed.

OUTPUT SPECIFICATIONS

Voltage Range:

0V to 60V

Current Range:

0A to 20A

Power Range:

Up to 420W

Output Voltage Setting:

By coarse and fine controls.

Output Current Setting:

By single logarithmic control.

Operating Mode:

Constant voltage or constant current with automatic cross-over
provided that the power demanded stays within the power envelope,
see graph. Outside of this envelope the output becomes unregulated.

XPF 60-20
POWER ENVELOPE

Output Switch:

Electronic. Preset voltage and current displayed when off.

Output Terminals:

4mm terminals on 19mm (0·75”) pitch. 30A max.

Sensing:

Switchable between local and remote. Spring-loaded push terminals
for remote connection.

Output Protection:

Forward protection by Over-Voltage Protection (OVP) trip; maximum
voltage that should be applied to the terminals is 70V.
Reverse protection by diode clamp for reverse currents up to 3A.

Over-temperature Protection: The output will be tripped off if a fault causes the internal

temperature to rise excessively.

OVP Range:

10% to 110% of maximum output voltage set by front panel
screwdriver adjustment.

Line Regulation:

Change in output for a 10% line change:

Constant voltage: <0.01% of maximum output

Constant current: <0.01% of maximum output

Load Regulation:

Change in output for any load change within PowerFlex envelope,
remote sense connected:

Constant voltage: <0.01% of maximum output

Constant current: <0.05% of maximum output

Ripple & Noise
(20MHz bandwidth):

4mVrms max; typically <2mVrms, <10mV pk-pk, at maximum load,
CV mode.