Maintenance, Storage – Xylem ISE Electrodes Ag-S 60 Benutzerhandbuch

Seite 17

Advertising
background image

Maintenance

x

For combination electrodes: replenish used reference

electrolyte solution.

x

In case of increased response times of the electrode

perform the following activities:

- Polish the membrane surface using an SF/K

polishing strip. To do so, moisten a polishing trip with

deionized water, and with circular movements polish

the membrane surface for approx. 30 seconds.

- Subsequently place the electrode into diluted

standard solution for approx. 5 minutes.

Storage

x

Between two measurements, place the combination

electrode into diluted standard solution.

x

Overnight up to one week: Push the closing ring
upwards to close the refilling opening. Then rinse
the combination electrode with deionized water, blot
it dry using a clean paper towel and store it in a dry
place.

x

For more than a week: Remove the electrolyte
solution from the reference electrode and rinse the
combination electrode with deionized water, blot it
dry using a clean paper towel and put on the
protection cap. Store the combination electrode in a
dry place.