Error diagnostics

Cause

Remedy

Measured value

unstable

Inner junctions not sufficiently

wetted with bridge electrolyte

Fill up bridge electrolyte until the

inner junctions are covered with

bridge electrolyte

Inner junctions encrusted

Leave the bridge electrolyte

to react on the inner junctions

for some hours until the

crusts have dissolved

Cable broken

Exchange cable

Slope too low

Membrane surface contaminated

Polish the membrane surface

with polishing strip S/FK (see

MAINTENANCE)

Conditioning time too short

Extend conditioning time

Standard solutions too old

Use new standard solutions

Inner junctions

encrusted

Leave the bridge electrolyte

to react on the inner junctions

for some hours until the

crusts have dissolved.

Electrode defective

Exchange electrode