Interferences

• Complexes with Ni

2+

, Cd

2+

, Cu

2+

• Strong reducing agents
• Interfering ions: 10 % error with the following ratios of

concentration:
(Ratio of concentration = Interfering ion / Measuring ion)

Aging

Please note that every electrode undergoes a natural

aging process. The response time increases and the
slope decreases with the age of the electrode. The
following factors shorten the lifetime considerably:

• Incorrect storing
• Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of

interfering ions)

• High temperatures
• High changes in temperature

The warranty does not cover damage caused by
measuring conditions and mechanical damage.