Maintenance

• For combination electrodes: replenish used reference

electrolyte solution.

In case of increased response times of the electrode

perform the following activities:

– Polish the membrane surface using an SF/K

polishing strip (see R

ECOMMENDED

ACCESSORIES

). To do so, moisten a polishing

trip with deionized water, and with circular
movements polish the membrane surface for
approx. 30 seconds.

– Subsequently place the electrode into diluted

standard solution for approx. 5 minutes.

Storage

Between two measurements, place the
combination

electrode into diluted standard

solution.

Overnight up to one week: Push the closing ring
upwards to close the refilling opening. Then rinse
the combination electrode with deionized water,
blot it dry using a clean paper towel and store it
in a dry place.

For more than a week: Remove the electrolyte
solution from the reference electrode and rinse
the combination electrode with deionized water,
blot it dry using a clean paper towel and put on
the protection cap. Store the combination
electrode in a dry place.