Xylem TA 197 - Oxi Oxygen sensor Benutzerhandbuch

Seite 26

Advertising
background image

Technical data

Empfohlenes Zubehör

Recommended accessories

26

27

Measuring principle

Membrane covered galvanic sensor

Temperature compensation

IMT

Meas. conditions

Meas. range

0 - 50 mg/l O2

Temperature range

0 °C - 50 °C

Max. admissible overpressure

10 bar

Immersion depth

min.

6 cm

max.

100 m depth

Operating position

any

Incident flow

> 3 cm/s at

10 % meas. accuracy

10 cm/s at

5 % meas. accuracy

18 cm/s at

1 % meas. accuracy

Characteristics when delivered

Zero signal

< 0.1 % from saturation value

Response time at 20 °C

t90 (90 % of indication of end value after) < 10 s
t95 (95 % of indication of end value after) < 16 s
t99 (99 % of indication of end value after) < 60 s

Self consumption at 20 °C

0.008 µg h-1 (mg/l)-1

Temperature adaptation

IMT-compensation

Drift

approx. 3 %/month in operation condition

Operation time

at least 6 months per electrolyte filling

Material

Membrane head

POM

Membrane

FEP

Sensor head

Epoxy, PEEK

Thermistor housing

VA-steel 1.4571

Shaft

POM

Connection head

VA-steel 1.4571

Dimensions

Shaft length

245 mm

Shaft diameter

38 mm

Membrane thickness

13 µm

Cable length

25 m to 100 m

Modell
Model

Best.-Nr.

O.-No.

Batterierührer

Battery stirrer

BR 325

203 826

Schutzkorb Stahl

Protection hood
steel

A 325/S/LF/SK

902 839

Schutzkorb Kunststoff

Protection hood
plastic

A 325/K/SK

903 834

Advertising