Xylem OxiTopIS 12-6 Benutzerhandbuch

Seite 31

Advertising
background image

Measurement

Disturbances / System messages

30

31

Recall stored values:
Press S until measured value is displayed (1 second).

Scroll to next day by repressing the S key while the measured
value is displayed (5 sec). Fast scrolling by repeatedly pressing
the S key.

· Convert the displayed measured value (digits) into the BOD value with the

following table (Digits x Factor = BOD5 in mg/l):

Sample volume (ml)

Measuring range (mg/l)

Factor

432

0 - 40

1

365

0 - 80

2

250

0 - 200

5

164

0 - 400

10

97

0 - 800

20

43.5

0 - 2000

50

22.7

0 - 4000

100

Disturbances

· Measured value remains below measuring range

The display shows zero or too low a value.
The measuring equipment is not water-tight.
Check rubber quiver, screw top and bottle.
Insufficient sample pretreatment or preservation.
The temperature of the sample had not sufficiently been adjusted (< 15°C).

· Measuring range exceeded.

The measuring range chosen is too small. With very high values (> 2000 mg/l) we
recommend to predilute the sample.
Nitrification inhibitor (allylthiourea) is missing or lacking.

Errors due to procedure have not been mentioned.

System messages

Memory empty (

IF = measured value of day 1 is missing)

Change batteries (approx. every 3 years)

Value remains below measuring range (< 0 digits)

Value exceeds measuring range (> 50 digits)