Cleaning / Battery replacement

Information / Technical data

32

33

Cleaning

Cleaning of sample bottles

Do not use disinfectants!
(Disinfectants will kill the required microorganisms!)

Remove

gross contaminations mechanically, e.g. with a brush.

Rinse the bottles with clear water or with water of the next sample.
(After using detergents rinse thoroughly! Detergent residues may disturb the BOD

5

determination!)

Cleaning of the OxiTop

â

Single Measuring System

Do not use alcohol or acetone!

Clean with a soft cloth and aqueous soap solution

Battery replacement (approx. every 3 years)

- Push in spring catch.
- Open upper lid.

- Remove batteries.
- Put in new batteries, pay attention to polarity!
(e.g. WTW-Modell Batt/Oxitop

в

в

в

в

, Order No. 209 012).

- Place upper lid with tap on locking pin (lower lid). Attention! Do not squeeze cable
connection.
- Close upper lid (make the hook click).

Disposal note:
Please dispose of the lithium batteries at a battery collection point.

Information

Never use sealing grease or other lubricants for your OxiTop

â

instruments. Some of these products contain solvents which cause
severe damages on the plastic housings.
The tightness of the BOD bottles is completely sufficient without grease.
Rough contaminations and particles on the contact surfaces of the
rubber quiver annd the OxiTop

â

must always be wiped off.

WTW grants no guarantee for damages caused by sealing grease.

Labelling:
To label the bottles clearly, WTW offers a marking set (6 marking rings with
numbers, which can be put on the neck of the bottle):

Model MARK-6 (6 pcs)
Order No. 209 013.

Check of the measuring system:
On request WTW offer special test instruments for test control.

Further literature:
WTW BOD handbook
DIN 38409T51
DIN 38409T52

Technical data OxiTop

в

в

в

в

single measuring system

Measuring principle

Respirometric (manometric)

Display

2-digit 7-segment LED, 10 mm

Measured parameter

Digits

Measuring range

0 ... 40 digits (+10 digits for overranging)

Admissible measuring
temperature

20°C

± 1 K

Admissible sample
temperature
(when being filled)

15-20°C

Power supply

Lithium batteries (260 mAh), 2 x type CR 2430 (3V)

Power consumption

Max. 25 mA (during measurement)

Protection class
Protection system

3 IEC 1010
IP 54 IEC 529

EMC

Emissions
Immunity

EN 50081-1, FCC Class A
EN 50082-2, NAMUR recommendation

Climate class

Ambient temperature

2, VDI/VDE 3540
Storage:

-25°C ... +65°C

Operation: +20°C

± 1K

Test mark

CE

Dimensions
Weight

H: 69 mm,

Æ

70 mm

Approx. 85 g