Gb 6. commissioning, operation – WIKA CPB3800 Benutzerhandbuch

Seite 30

Advertising
background image

30

WIKA operating instructions model CPB3800

14062232.03 12/2013 GB/D

GB

6. Commissioning, operation

6.6.2 During calibration

When the pressure balance is correctly set up and there are no leaks the piston

should "float" for many minutes without it being necessary to touch the spindle pump

handwheel. On the initial setting up, however, there may be some air trapped in the base

of the piston/cylinder unit. As this leaks past the piston the weights may fall slightly but it

will only be for a matter of a few minutes until the air has escaped. If the piston continues

to fall, check the connections for leaks.

During calibration, the weights should be rotated by hand. It is desirable that the weights

should only be rotated when approximately the correct pressure is obtained. Weights

should not be brought to rest by fully releasing the pressure and allowing the piston

head to rotate against its stop under the full load of the weight pile.

It is essential that the weights spin freely during readings. The piston stops moving when

the pressure is too high or too low. At the lowest pressures the weights will not spin

for more than a few seconds unless a very thin oil is used, but providing the weight is

rotated by hand before taking a reading and is obviously "floating" an accurate reading

will be given.

CAUTION!

Care should be exercised at all times when rotating the weights. Failure

to do so may cause damage to the actual piston unit, or possibly injury to

the operator.

Therefore, the rotational motion should be stopped by hand. Only then

new masses for further test points can be placed or the pressure can be

released completely.

6.6.3 Datum levels

When testing gauges on liquid it is occasionally necessary to take into account heads

of liquid since a height difference of 10 mm corresponds to approximately 1 mbar. The

datum levels of the models CPB3800 piston units are marked with a groove on the

outer diameter of the piston unit. It should be noted that when the pressure balance is

re-calibrated by a laboratory other than DH-Budenberg/WIKA, the datum level at which

the tests have been carried out may differ from this standard and therefore allowance

should be made for any variation.

The drawing shows the head effect that may have to be compensated for when high

accuracy calibration is desired. The following formula will enable the head correction to

be calculated.

Advertising