WIKA CPH65I0 Benutzerhandbuch

Seite 21

Advertising
background image

21

WIKA operating instructions, model CPH65I0

14061681.02 03/2013 GB/D

GB

6. Commissioning, operation

6. Commissioning, operation

As soon as the model CPH65I0 hand-held pressure calibrator is turned on, by pressing

the ON/OFF key, it will go through a short self-test routine. During this routine, the

display shows the current firmware version, auto-shutdown status and the measuring

range of the internal pressure sensor. The calibrator requires a maximum of 5 minutes

warm-up to reach its specified accuracy. Large changes in ambient temperature may

make a longer warm-up period necessary. The pressure display of the calibrator should

be zeroed before starting the pressure calibration. See chapter 6.3 "Use of the "ZERO"

function".

6.1 Calibrator display

The calibrator's display consists of two sectors: The menu bar (located along the

bottom of the screen) is used to access the menu system. The main display (the rest of

the display) consists of up to three sub-sectors for the measurement activities. These

sub-sectors will be referred to as the UPPER, MIDDLE and LOWER displays in this

document.

Figure 1 shows the location of the different display fields, which are described in the

following table.

1) Pressure units

Indication of the pressure unit (selectable from 16 pressure units)

2) Units

Indication of the measuring unit

3) Display of the span

Indication of 4 ... 20 mA span (only with mA measurement)

4) Menu list
5) Primary parameters

Indication of the current measured parameters

Figure 1: Display

1

2

3

4

5