Gb 6. commissioning, operation – WIKA CPH65I0 Benutzerhandbuch

Seite 23

Advertising
background image

23

WIKA operating instructions, model CPH65I0

14061681.02 03/2013 GB/D

GB

6. Commissioning, operation

The following table shows which functions are available concurrently.

An "--" in a column indicates that the mode in the active display will not be available for

selection if the mode in that row is in use in any other display sector.

Active display

Other displa

ys

P[1]

P[2]

P[1] ST

P[2] ST

mA

RTD

P[1]

P[2]

P[1] ST

--

--

--

P[2] ST

--

--

--

mA

--

--

RTD

6.1.1.3 Accessing other menus

In order to access further menus, select the "MORE" option in the main menu. See

chapter 6.6 "Menu structure".

6.2 Using the backlighting

The backlight has its own dedicated button. The light is toggled on and off when the

button is pressed.

6.3 Using the "ZERO" function

When the ZERO key is pressed, the calibrator will zero the {active display}, as long as

a pressure mode is selected and the pressure is within the zero limit. The zero limits are

within 5 % of the full scale range of the selected sensor. If the display indicates "OL" the

zero function will not operate.

The ZERO key is only used for pressure.

Advertising