WIKA operating instructions portable SF

6

gas transfer unit, model GTU-10

14051107.01 06/2013 GB/D

GB

6. Commissioning, operation

CAUTION!

Danger of tripping caused by improper routing of machine connec-

tions

Improperly installed machine connections such as cables, hoses or

pipelines can create tripping hazards, and can cause serious injury.

Route the machine connections so that no tripping sources are

present.

Do not use machine connections to assist carrying.

Mark unavoidable tripping sources with yellow and black marker tape.

Location

To prevent the risk of falling, work in an environment in accordance with EN12464 with

sufficient lighting, ≥ 300 lx.

To ensure adequate ventilation of the instrument, the outlet of the fan must not be

obstructed. Ensure a min. clearance of 30 cm.

Mounting and disassembly of components

All hoses or fittings and threaded connections connected to the GTU-10 must be

designed for operation of at least 50 bar.

Please be aware that the hose lines may be under high pressure and should be vented

into a tank after use.

The connections are sealed with plastic protective caps or adhesive foil. These protec-

tive caps and the adhesive foils must be removed before commissioning.

Protective system

The housing must, at all times, be closed and intact. Any operation with open or defec-

tive housing will void the operating licence.

Electrical connection

The power supply for the GTU-10 is made via male rubber connector in accordance to

IEC 60320 / C14. The permissible voltage range is indicated on the rating plate. Should

the customer's own power cable be used, it must have a minimum cross-section of

1.5 mm

2

.

The unit is only suitable for use in industrial environments.

There should be no connection to the public low-voltage supply without permission from

the power supply company concerned.

Behaviour on power loss

If the power should fail during operation, the electrical control system will prevent the

automatic restart of the compressor.