Instructions for use

1.2.1 Dwell time
For the cool packs 0.5 ml or 1.5/2.0 ml the dwell time (IsoTherm time) has been

calculated for 3 use cases each (see measurement in the table):

Insulating box with cover (a)

Open insulating box (b)

Cool pack and tube holder on the tube holder cover (c)

¹ Framework conditions for measuring the IsoTherm time:

Maximum filling of the tube holder.

Micro test tubes filled to the maximum.

Environment at room temperature, system at rest.

Measurement of the sample temperature.

² For IsoPack 1.5/2.0 ml: only valid for central bore holes.

Use

case¹

a

b

c

Pack

version

0 °C to

4 °C

(white)

-21 °C to

18 °C

(blue)

0 °C to

4 °C

(white)

-21 °C to

-18 °C

(blue)

0 °C to

4 °C

(white)

-21 °C to

-18 °C

(blue)

IsoTherm

time

>6 h

>3 h

>4 h

>1 to 2 h

> 1.5 to 3 h > 0.5 to 1 h²

Sample temperature

Cold pack 0 °C

Cold pack –21 °C

T (°C)

t (h)

+12

+8

+4

0

0

1

2

3

4

5

6

7

8

9

–4

–8

–12

–16

–20

–24